Cook Medical Zenith Alpha Abdominal Endovascular Graft (ICN 1112)
Important Customer Notice Recall
Ref: 2020/1112
ICN Number: 1112
MHRA 2020/008/010/487/007
NHSSC: 38
Alert:
- Cook Medical has identified that specific lots of the Zenith Alpha Abdominal Endovascular Graft (ZIMB main body) may contain excess glue within the handle of the delivery system.
- This may result in an inability to fully deploy the graft via the standard method or troubleshooting deployment method provided in the IFU.
- Potential adverse events that may occur if an affected product is used include a prolonged procedure and/or open surgical repair.
Products Affected:
- There are 76 affected products available via NHS Supply Chain’s eDirect route.
- See our Downloads ▼ section to refer to the Field Safety Notice and product listing for further information on the affected product codes.
- Please note only the lot numbers listed in the Field Safety Notice are affected.
- Devices already implanted are not affect by this recall.
Next Steps:
- Read and follow the full instructions in the Field Safety Notice and share this document with all users of the affected products.
- Quarantine any affected product/ lots.
- Customers are to complete the Customer Reply Form and email it to European.FieldAction@CookMedical.com.
- If you have any of the affected products please contact your NHS Customer Service Advisor with full details to enable us to monitor a return direct to the supplier and manage your credit.
If you have any further questions, please contact your local NHS Supply Chain Customer Service Advisor.