Intersurgical Adult EcoliteTM Aerosol Mask Additional Lots (ICN 1169)

Important Customer Notice Recall

Ref: 2020/1169 ICN Number: 1169

NHSSC : 63A

Update

  • Intersurgical has issued an updated Field Safety Notice 301083-A which supersedes 301083.   A further four lot numbers have been added, bringing the total to be recalled to seven:
NPC Product Description Supplier Code/MPC Additional Affected Lots February 2021
FDD4082 Cirrus nebuliser Adult Eco mask kit Ecolite mask kit and tube 1453015 32052758,32053066,32053274,32053357,
  • Read and follow the instructions in this latest Field Safety Notice. See our Downloads ▼ section
  • Where affected lot numbers are identified, please quarantine and contact your NHS Supply Chain Customer Service Advisor to arrange a return for full credit (reason code 7 Supplier Recall). All products to be returned no later than Friday 5 March 2021.

NHSSC : 63

Alert: 

  • Intersurgical has issued a Field Safety Notice for specific lots of Cirrus nebuliser Adult Eco mask kit Ecolite mask kit and tube.
  • Intersurgical received a complaint where a product that should contain an Adult EcoliteTM Aerosol Mask with vents, has been found to contain a similar mask without vents.
  • This would prevent the patient breathing normally and could result in distress or some form of harm to the patient if unnoticed.

Products Affected:  

  • There is one NHS Supply Chain product code affected.
NPC Product Description Supplier Code/MPC Affected Lots
FDD4082 Cirrus nebuliser Adult Eco mask kit Ecolite mask kit and tube 1453015 32052973,32053276,32053524
  • Please see our Downloads ▼ section to refer to the Field Safety Notice for further information on the affected product and specific lot numbers.

Next Steps:

  • Read and follow the full instructions in the Field Safety Notice and share this document with all users of the affected products.
  • Where affected lot numbers are identified, please contact your NHS Supply Chain Customer Service Advisor to arrange a return for full credit (reason code 7 Supplier Recall). All products to be returned no later than Friday 14 January 2021.

If you have any further questions, please contact your local NHS Supply Chain Customer Service Advisor.

Downloads ▼

  • 1169 Intersurgical Field Safety Notice

    A PDF document of the Field Safety Notice for ICN 1169.
  • 1169 Intersurgical Update Field Safety Notice

    A PDF of the updated Field Safety Notice for ICN 1169.