Skip to main content

Field Safety Notice Baxter Healthcare Limited Welch Allyn Power Cords FFF3068 (ICN 2572)

Important Customer Notice Update

Ref: 2024/2572 ICN Number: 2572

  • Baxter has issued an updated Field Safety Correction Notice for the power cords used with the Welch Allyn ProBP 3400 Digital Blood Pressure Device and Welch Allyn Spot Vision Screener following reports of an issue relating to the cord construction not meeting the insulation rating for international standards.
  • Baxter has improved the power cord’s insulation and is asking customers to submit a request for replacement power cords.
  • Please submit your request for replacement power cords by contacting Baxter Customer Support at HRC_ordersUK@baxter.com.
  • Healthcare providers may continue using the affected power cords while waiting for replacement power cords. If fraying or damage is observed, users should avoid contact with the cord and should discard the power cord immediately.
  • Once you receive replacement power cords, discard any non-compliant power cords.

Download our publications ▼

  • 2572 Baxter Field Safety Notice Update 27 November 2024

    A PDF showing the Field Safety Notice Update for ICN 2572.
  • 2572 Baxter Field Safety Notice 17 June 2024