Field Safety Notice Becton Dickinson BD® Connecta™ Stopcock FVK011 (ICN 3551)

Important Customer Notice Status New Type Product Recall

Ref: 2026/3551 ICN Number: 3551

NHSSC: 88

Alert:

  • Becton Dickinson has issued Field Safety Notice MDS-26-06177 having identified specific Lot numbers for two of its BD® Connecta™ stopcocks which may have been manufactured with an inadequate amount of solvent.
  • The lack of adequate solvent may result in separation of the extension tubing from the male luer component.
  • Separation of the Connecta™ stopcock extension tubing may occur during use, resulting in a compromised fluid pathway. This may result in leakage with potential consequences including under-infusion or disruption of prescribed therapy, exposure to potentially hazardous fluids, infection or air embolism.
  • The separation may be apparent upon initiation of infusion, or during standard patient care at which point the healthcare professional can promptly discontinue use and replace the affected stopcock extension set.

Products Affected:

  • Two products are affected.
  • One of products affected is available via NHS Supply Chain:
NPC Product Description Supplier Code / MPC

Route

FVK011

Stopcock with extension line 3 way with 10cm extension line total length including stopcock 16cm 394995

Stocked

  • See our Downloads ▼ section to access the product listing for further information on the affected product codes and potential alternatives and two view an extract from the supplier’s website of affected Lot numbers.
  • Please note only the Lot numbers quoted in the Field Safety Notice are affected.
  • All NHS Supply Chain stock has been checked, and all future issues will be of unaffected, different Lot numbers.
  • Becton Dickinson has confirmed that all other active Connecta™ stopcock product codes are not affected by this issue and there are currently no expected disruptions to future supply.

Next Steps:

  • Read and follow the full instructions in the Field Safety Notice and share with all users of the affected products.
  • Cease use and quarantine all affected Lot numbers.
  • If the product is currently in use, replace immediately with unaffected product.
  • If the product is already in use and you are unable to determine if it is within scope as the packaging or other records are not available, it is recommended to replace with unaffected product.
  • If an affected product has already been used and is no longer on site, no patient follow up activities are required.
  • Becton Dickinson has advised that, during its investigation, some reported cases involved use with power injection. Clinical users are reminded of the warning stated in the Instructions for Use:
    • DO NOT use with power injectors. Doing so may result in leakage and/or failure of the device.
  • Contact your local NHS Supply Chain Customer Services Advisor by Friday 11 December 2026 with details of any affected products you hold, including the National Product Code (NPC), Lot number(s), and quantity. Your Customer Services Advisor will give you a unique CMS log number.
  • If completing the customer response form provided by Becton Dickinson, ensure you quote your full details and email to: Customer Service Admin Support coreadminsupport@supplychain.nhs.uk.
  • Dispose of the affected products in accordance with your local regulations. Upon written confirmation of disposal in the form of an email, quoting the CMS reference, your Customer Services Advisor will raise the appropriate credit.
  • Where potential alternatives are identified, customers are advised to consult their own clinical experts to ensure suitability for their organisation’s use of these products.

If you have any further questions, please contact your local NHS Supply Chain Customer Service Advisor or the Acute team.

Please be aware that in the event of a Field Safety Notice/Product Recall, we may need to provide manufacturers, UK responsible Persons, and Distributers with contact details of customers who have potentially received the affected stock. This is to help them to reconcile their stock and evidence to the regulators that all actions have been taken to ensure that the unused products have been removed from customers to prevent inadvertent use of faulty and potentially unsafe products.

Download our publications ▼

  • 3551 Becton Dickinson Field Safety Notice 17 November 2026

    A PDF showing the Field Safety Notice for ICN 3551.
  • 3551 Becton Dickinson Product Listing with Alternatives 17 November 2026

    An Excel file showing the product listing with alternatives for ICN 3551.
  • 3551 Becton Dickinson Affected Lot Numbers 17 November 2026

    An Excel file showing the affected lot numbers for ICN 3551.
  • 3551 Becton Dickinson Response Form 17 November 2026

    A PDF showing the customer response form for ICN 3551.