Field Safety Notice Medtronic Abre™ Venous Self-Expanding Stent System (ICN 3557)
Important Customer Notice Status New Type Product Alert
NHSSC: 91
Alert:
- Medtronic has issued Field Safety Corrective Action Notice FA00007 to inform all users of a forthcoming update to the Instructions for Use (IFU) for the Abre™ venous self-expanding stent system.
- This update provides a validated manual deployment workaround method to help mitigate potential harms in the rare event of a partial stent deployment.
- Medtronic has advised it has completed validation testing to standardise the deployment workaround method, as shown in pages 2,3 and 4 of this notice, and that the Abre’s IFU will be updated to reflect this standardised approach, pending regulatory approvals.
- These instructions should only be followed in the event a partial stent deployment occurs.
- Medtronic is taking this action to reduce the severity of potential patient harms.
- The steps provided on the Field Safety Notice are intended to minimise the risk of serious adverse health consequences in the rare event of partial stent deployment.
Products Affected:
- There are 31 products affected.
- All of the products affected are available via NHS Supply Chain.
- See our Downloads section to access the product listing for further information on the affected product codes.
- Please note that this is an update to the Instructions for Use, no product withdrawal or return is required.
Next Steps:
- Read, share and follow the full instructions in the Field Safety Corrective Action Notice.
- Complete the customer acknowledgement form and return to fsnuki@medtronic.com.
- If you have any questions regarding this issue or communication, please contact your Medtronic Field Representative.
- Medtronic has confirmed that it has already contacted all customers affected by this notice.
If you have any further questions, please contact your local NHS Supply Chain Customer Service Advisor or the Acute team.
Please be aware that in the event of a Field Safety Notice/Product Recall, we may need to provide manufacturers, UK responsible Persons, and Distributers with contact details of customers who have potentially received the affected stock. This is to help them to reconcile their stock and evidence to the regulators that all actions have been taken to ensure that the unused products have been removed from customers to prevent inadvertent use of faulty and potentially unsafe products.