Supply Issues Nova Biomedical StatStrip® Ketone Test Strips (ICN 3503)

Important Customer Notice Status Update Type Supply Disruption

Ref: 2026/3503 ICN Number: 3503

  • Following Nova Biomedical’s Field Safety Notice to notify StatStrip® Glucose / β-Ketone hospital meter system customers of specific Lots of StatStrip® ketone test strips that have demonstrated a higher-than-expected rate of failure during quality control (QC) testing, Nova Medical has issued customer bulletin 07-26SS-UK to advise of a temporary change to the product’s shelf life.
  • As part of Nova Biomedical’s ongoing commitment to product quality and customer support, and to address the impact of a shortage in supply of StatStrip® ketone test strips globally, Nova has implemented a process improvement to manufacture and release five new Lots of ketone test strips demonstrating reliable QC test performance.
  • Nova is repeating stability studies for these Lots of test strips which will have a reduced shelf life.
  • Previously, StatStrip® ketone test strips had a 24-month shelf life. The five new Lots of test strips that have been manufactured and are ready for shipment will have a 2-month (60-day) shelf life and are listed in the table below:

Part Number

Description Lot Number Date of Manufacture

Expiration Date

P/N 46951

TEST STRIP 50CT KET STRIP DUAL PACK 2226280029 7 October 2026 7 December 2026

2226281029

8 October 2026

8 December 2026

2226286029 13 October 2026

13 December 2026

2226287029

14 October 2026

14 December 2026

2226288029

15 October 2026

15 December 2026

  • This current 2-month shelf-life is supported by real-time stability data.
  • Nova will continue to test and verify strip stability and QC testing performance and to ensure that all Lots of strips meet our quality standards.
  • Nova will continue to manufacture additional Lots to ensure that supply-demand, patient testing needs, and regulatory requirements are met.
  • See our Downloads section to view the supplier’s letter for more information on StatStrip® reduced shelf-life notification.

Download our publications ▼

  • 3503 Nova Biomedical Supplier Letter 6 November 2026

    A PDF showing the supplier's letter for ICN 3503.
  • 3503 Nova Biomedical Patient Safety Alert 14 October 2026

    A PDF showing the patient safety alert for ICN 3503.
  • 3503 Nova Biomedical Product Listing 16 October 2026

    An Excel file showing the product listing for ICN 3503.