Supply Issues Nova Biomedical StatStrip® Ketone Test Strips (ICN 3503)
Important Customer Notice Status Update Type Supply Disruption
Ref: 2026/3503
ICN Number: 3503
- Following Nova Biomedical’s Field Safety Notice to notify StatStrip® Glucose / β-Ketone hospital meter system customers of specific Lots of StatStrip® ketone test strips that have demonstrated a higher-than-expected rate of failure during quality control (QC) testing, Nova Medical has issued customer bulletin 07-26SS-UK to advise of a temporary change to the product’s shelf life.
- As part of Nova Biomedical’s ongoing commitment to product quality and customer support, and to address the impact of a shortage in supply of StatStrip® ketone test strips globally, Nova has implemented a process improvement to manufacture and release five new Lots of ketone test strips demonstrating reliable QC test performance.
- Nova is repeating stability studies for these Lots of test strips which will have a reduced shelf life.
- Previously, StatStrip® ketone test strips had a 24-month shelf life. The five new Lots of test strips that have been manufactured and are ready for shipment will have a 2-month (60-day) shelf life and are listed in the table below:
|
Part Number |
Description | Lot Number | Date of Manufacture |
Expiration Date |
|
P/N 46951 |
TEST STRIP 50CT KET STRIP DUAL PACK | 2226280029 | 7 October 2026 | 7 December 2026 |
|
2226281029 |
8 October 2026 |
8 December 2026 |
||
| 2226286029 | 13 October 2026 |
13 December 2026 |
||
|
2226287029 |
14 October 2026 |
14 December 2026 |
||
|
2226288029 |
15 October 2026 |
15 December 2026 |
- This current 2-month shelf-life is supported by real-time stability data.
- Nova will continue to test and verify strip stability and QC testing performance and to ensure that all Lots of strips meet our quality standards.
- Nova will continue to manufacture additional Lots to ensure that supply-demand, patient testing needs, and regulatory requirements are met.
- See our Downloads section to view the supplier’s letter for more information on StatStrip® reduced shelf-life notification.
- The MHRA has issued an update to Device Safety Information (DSI/2026/013) regarding an additional Lot of Nova Biomedical StatStrip ketone test strips that is affected:
- 2226117249.
- See our Useful Links section to view Field Safety Notice ICN 3500 for more information on the product recall and to view DSI from the MHRA.
- To support customers to identify and validate alternative ketone testing methods, including alternative point-of-care devices and laboratory testing, we have complied an updated product listing.
- See our Downloads▼ section to access the product listing for further information on the affected product codes, point-of-care testing suppliers, and their ketone products including meters, strips and QC solutions available to order through our catalogue.
- Please note customers using Abbott Diabetes products should contact Abbott Diabetes directly as these products are not available via NHS Supply Chain.
- The Department of Health and Social Care has issued a National Patient Safety Alert, NatPSA/2026/007/NSDR, regarding the supply disruption for Nova Biomedical StatStrip® blood ketone test strips.
- See our Downloads▼ section to access the National Patient Safety Alert.
- The alert requires actions to be completed by 20 October 2026.
- Organisations should identify a clinical lead and form a working group to act on the Field Safety Notice issued by Nova and follow the advice issued by MHRA in its device safety notice.
- This will include taking action to:
- Discontinue clinical use and dispose of all Lots identified by Nova as subject to recall.
- Assess current and anticipated blood ketone testing capacity, including the number and location of ketone testing devices, current usable strip stock and anticipated out-of-stock dates.
- Identify and validate alternative ketone testing methods where available, including alternative point-of-care devices and laboratory testing.
- Prioritise available blood ketone testing for patients at greatest clinical risk where supply is constrained.
Change:
- The supplier Nova Biomedical has informed us of a supply issue with its StatStrip® ketone test strips.
- One product is affected:
- Supplier Code: 46951 – Blood Ketone Test Strips For use with StatStrip Meters.
- The affected product is supplied through seven NHS Supply Chain product codes due to regional grouping:
- HHD851 − Ketone test strips for use with StatStrip Meters
- HHD85007 − Ketone test strips for use with StatStrip Meters
- HHD412 − Ketone test strips for use with StatStrip Meters
- HHD722 − Ketone test strips for use with StatStrip Meters
- HHD835 − Ketone test strips for use with StatStrip Meters
- HHD839 − Ketone test strips for use with StatStrip Meters
- HHD553 – Ketone test strips for use with StatStrip Meters.
- Please note that HHD851 is available to all customers.
- All other product codes are regionally available and not visible to all customers.
- The supply issue is due to the supplier issuing a Field Safety Notice having identified quality issues affecting specified Lots of StatStrip® ketone test strips.
- The MHRA has issued Device Safety Information (DSI/2026/013) regarding the recall and supply of the affected product.
- See our Useful Links section to view Field Safety Notice ICN 3500 for more information on the product recall and to view DSI from the MHRA.
Impact:
- Nova Biomedical has paused distribution of the affected Lots and advised that any further Lots found to fail quality-control (QC) testing should also be withdrawn from use.
- The recall, paused distribution, and potential withdrawal of additional Lots may result in a progressive reduction in the availability of ketone test strips and disruption to blood ketone point-of-care testing.
- NHS England has issued Interim clinical prioritisation guidance for blood ketone testing during the supply disruption.
- See our Useful Links section to view the interim clinical prioritisation guidance issued by NHS England.
- The affected product codes are subject to intermittent supply.
- Unfortunately, there are no alternatives available.
- See our Downloads ▼ section to access the product listing for further information on the affected product codes.
- Nova is currently expecting new product to be available during the week commencing 26 October 2026.
- The supply issue is anticipated to be resolved in approximately four weeks.
Next Steps:
- Customers are advised to log into our website to view the product listing for further information on the affected product codes.
- Should you have any questions please contact Nova Biomedical UK:
- Telephone: +44 1928 704047
- Email: EU-QA@novabio.com.
- NHS Supply Chain is in daily contact with Nova, Department of Health and Social Care (DHSC) and NHS England and we will update this ICN as soon as we have further information.
- If you have any further questions, please contact your local NHS Supply Chain Customer Services Advisor.
Please note that due to the clinical nature of these alternative products, we advise you to consult your own clinical experts to ensure suitability for your organisation’s use of these products.