Field Safety Notice B Braun Infusomat Space Transfusion Line – Leakage (ICN 1991)

Important Customer Notice Field Safety Notice

Ref: 2023/1991 ICN Number: 1991

Alert:

  • B Braun has issued a Field Safety Corrective Action Notice after identifying the risk for leakages in four batch numbers of their Infusomat Space Transfusion Lines.
  • The potential leakage is located between the tube and the patients’ Luer connector.
  • B Braun advise no serious injuries to patients, users, or third parties have been reported to date, however the deviation might harbour the risk for the patient of microbial contamination, under supply, open patient access, or air infusion.

Products Affected:

  •  The product affected is available from NHS Supply Chain:
NPC Product Description Supplier Code / MPC  Batch Numbers
FSB2495 Administration set for Infusomat Space Volumatic Pump 8270066SP-01 22F19E8ST5

22G01E8ST5

22G03E8ST5

22H25E8ST5

 

  • Please note only the batch numbers quoted are affected. NHS Supply Chain Warehouses are being checked for these and future deliveries should be of different, unaffected batch numbers.
  • B Braun advise this withdrawal should have no impact on future supplies.

Next Steps:

  • Read and follow the full instructions in the Field Corrective Notice and share with all users of the affected products.
  • Cease use and quarantine affected bath numbers. It is recommended to exchange devices of these batch numbers which are currently in use.
  • Contact your local NHS Supply Chain Customer Service Advisor with full details of affected, unused, products for return for full credit using reason code 7 Supplier recall. All devices to be returned no later than Friday 3 February 2023.

Downloads ▼

  • 1991 B Braun Field Safety Notice 16 January 2023

    A PDF showing the Field Safety Notice for ICN 1991.